In this research are used two types of detectors scintillation NaI(Tl) sizes (2"x2") ,(1"x1.5")for a
comparative study between them and are used source of radioactive cesium has energy (0.662 MeV) and
calculating the total area of the spectrum space scattering and area of peak optical and portability energy
analysis, found the size large for crystal detector leads to probability escape photons outside the crystals to be
less because it can reveal again and recorded pulse with pulse recorded by the electron apostate so increase the
value of portability energy analysis, and increase the volume of detectors cause increased probability for
scattering Compton ,In other words the increase in the probability of interaction effect photoelectric be less than
the increase in the probability of interaction scattering Compton so the Net Area under the peak at less an
increase of scattering spectrum.
تاريخ النشر
01/03/2014
الناشر
Journal of Information Engineering and Applications